Japan Zirconia Oxygen Analyzer ZR402G High Accuracy Genuine Original Product Price Negotiable

This analyzer is most suitable for monitoring the oxygen concentration of combustion gases in large or small boilers,. various industrial furnace and combustion The separate type converter ZR402G incorporates a LCD touch-screen for ease of operation. ... The integrated type ZR202S integrates both probe and converter Yokogawa's zirconia oxygen analyzers are used for combustion monitoring and control in a wide variety of applications. Models: ZR22G, ZR22S, ZR402G, ZR22P,
Oct 29th,2026 1 Views
  • The separate type converter ZR402G incorporates a LCD touchscreen for ease of operation.
  • The built-in heater assembly of the probe can be replaced on site, reducing maintenance costs (ZR22G only).
  • The probe uses a long-life, high-reliability, field replacable Zirconia sensor.
  • Remote maintenance using digital communications (HART) reduces maintenance costs.

Note: HART is a registered trademark of HART Communication Foundation.

Measurement object Oxygen concentration in combustion exhaust gas and mixed gas
Measurement system Zirconium oxide sensor
Measurement range 0.01 to 100 vol% O2
Output signal 4 to 20 mA DC, two points (ZR402G) (maximum load resistance 550Ω)
Setting range Any setting in the range of 0 - 5 vol%O2 to 0 - 100 vol%O2 (in 1 vol% O2), or partial range
Digital communication (HART) 250 to 550Ω, depending on number of field devices connected to the loop (multi-drop mode)
Sample gas pressure -5 to +250 kPa (Pressure compensation option necessary over +5 kPa)
Sample gas temperature 0 to 700℃ (Probe only)
700 to 1871℃ (with 0.15m probe with JIS 5K 32 FF flange and ZR22P High Temperature Probe Adapter)
Probe length 0.15, 0.4, 0.7, 1.0, 1.5, 2.0, 2.5, 3.0, 3.6, 4.2, 4.8, 5.4 m
Contact output Four points (one is fail-safe, normally open)
Contact input Two points
Calibration Method; zero/span calibration
Calibration mode; automatic, semi-automatic and manual (with an LCD touchscreen)
Construction of detector (ZR22G) Heater and thermocouple replaceable construction. 
Non-explosion-proof. Equivalent to NEMA4X/IP66
Construction of converter (ZR402G) Non-explosion-proof, outdoor installation, equivalent to NEMA4X/IP66
Ambient Temperature Probe (ZR22G): -20 to 150℃
Converter (ZR402G): -20 to 55℃)
Reference Gas System

Instrument Air (preferred): 50 kPa (200 kPa with check valve) + the pressure inside the furnace; Approx. 1 Nl/min consumption

Natural convection

Power supply 85 to 264 V AC; 45 to 66 Hz
Measurement object Oxygen concentration in combustion exhaust gas and mixed gas
Measurement system Zirconium oxide sensor
Explosion-proof Approval ZR22S-A (ATEX); EEx d IIB + H2 T2 Gb, Ex tb IIIC T300°C Db
ZR22S-B (FM); Class I, Division1, Groups B, C and D
ZR22S-C (CSA); Class I, Division1, Groups B, C and D
ZR22S-D (IECEx); Ex d IIB + H2 T2 Gb, Ex tb IIIC T300°C Db
ZR22S-K (KOSHA); Ex d IIB + H2 T2
Measurement range 0.01 to 100 vol% O2
Output signal 4 to 20 mA DC, two points (ZR402G) (maximum load resistance 550Ω)
Setting range Any setting in the range of 0 - 5 vol%O2 to 0 - 100 vol%O2 (in 1 vol% O2), or partial range
Digital communication (HART) 250 to 550Ω, depending on number of field devices connected to the loop (multi-drop mode)
Sample gas pressure ±5kPa
For 0.15m probe, -0.5 to +5 kPa.
No pressure fluctuation in the furnace should be allowed
Sample gas temperature 0 to 700℃
700 to 1871℃ (with 0.15m probe with JIS 5K 32 FF flange and ZR22P High Temperature Probe Adapter (certain options only))
Probe length 0.15, 0.4, 0.7, 1.0, 1.5, 2.0 m
Contact output Four points (one is fail-safe, normally open)
Contact input Two points
Calibration Method; zero/span calibration
Calibration mode; automatic, semi-automatic and manual (with an LCD touchscreen)
Construction of detector (ZR22S) Heater and thermocouple replaceable construction, equivalent to NEMA4X/IP66
Construction of converter (ZR402G) Non-explosion-proof, outdoor installation, equivalent to NEMA4X/IP66
Ambient Temperature -20 to +60℃ (-20 to +150℃ on the terminal box surface) (ZR22S)
-20 to +55℃ (ZR402G)
Reference Gas System Instrument Air: 50 kPa + the pressure inside the furnace; Approx. 1Nl/min consumption
Wiring Connection ATEX; M20 by 1.5 mm or 1/2 NPT select one type (4 pieces)
FM; 1/2 NPT (4 pieces)
CSA; 1/2 NPT (4 pieces)
IECEx; M20 by 1.5 mm or 1/2 NPT select one type (4 pieces)
Power supply 85 to 264 V AC; 45 to 66 Hz